Test frequency for a target depth
Frequency is the variable the operator controls directly, and it sets the standard depth of penetration through delta = 1 / sqrt(pi f mu sigma). Rearranged for frequency this becomes f = rho / (pi mu0 mur delta^2). The relationship is inverse-square: quadrupling the frequency halves the depth, so frequency selection is coarse and a factor of two either way is a real change in what the test can see.
Rather than asking for a depth directly, working practice asks how many standard depths deep the target lies. At n standard depths the eddy current field has fallen to e^-n of its surface value and lags the surface current by 57.3 x n degrees. Setting n = 1 puts the target at 37 % field and 57 degrees of lag, which is the usual compromise between penetration and signal. Setting n near 2 or 3 buys depth at the cost of amplitude (13 % and 5 %) and pushes the phase lag towards 115 and 172 degrees.
Beyond roughly three standard depths the response is unusable rather than merely weak: the phase has rotated so far that a deep flaw plots close to the lift-off direction and can be mistaken for probe wobble. If the wall is thicker than about three standard depths at the frequency you need for sensitivity, the answer is a different method or a dual-frequency approach, not more gain.
What comes out of this calculation is a starting frequency. Confirm it on the calibration standard: the frequency that gives the required signal-to-noise and the required phase separation between the shallowest and deepest calibration reflectors is the frequency the procedure gets.
Worked example
| Material property entered as | rho |
| Electrical resistivity | 72 uohm.cm |
| Conductivity | 2.39 %IACS |
| Relative permeability | 1 |
| Target depth | 1.5 mm |
| Standard depths at the target | 1 |
| Required standard depth | 1.5 mm |
| Test frequency | 81.06 kHz |
| Phase lag at target depth | 57.3 deg |
| Field strength at target depth | 36.8 % |
304 stainless (rho = 72 uohm.cm, mur = 1), 1.5 mm wall, one standard depth at the far surface. Required delta = 1.5/1 = 1.5 mm. In SI: rho = 7.2e-7 ohm.m, delta = 1.5e-3 m, delta^2 = 2.25e-6 m^2, pi.mu0 = 3.947842e-6. f = 7.2e-7 / (3.947842e-6 x 2.25e-6) = 7.2e-7 / 8.882644e-12 = 81057 Hz = 81.06 kHz. Cross-check with the engineering form 2533 x 72 / 2.25 = 81057 Hz. Phase lag at one standard depth = 1 radian = 57.3 degrees; field = 100 x e^-1 = 36.8 %.
Use at your own risk — verify before you act
These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.