Eddy current phase lag with depth
Phase lag is what turns eddy current testing into a depth-measuring method instead of an amplitude-measuring one. The current at depth x lags the surface current by x/delta radians, which is 57.3 degrees for every standard depth of penetration. Amplitude decays over the same distance, but amplitude is also driven by flaw volume, probe wobble and lift-off, whereas the phase angle is dominated by depth alone. That is why depth is sized from the angle of the signal on the impedance plane and not from its height.
On the impedance plane this is the whole basis of mixing and discrimination. With the display rotated so that lift-off lies horizontal, a surface-breaking flaw at the near surface plots almost perpendicular to lift-off, and as the flaw origin moves away from the coil the signal rotates progressively clockwise. Two reflectors that differ in depth by half a standard depth are separated by about 29 degrees – comfortably resolvable. Two that differ by a tenth are separated by 5.7 degrees and will not be separable in practice.
The rotation is also the practical depth limit. At three standard depths the lag is 172 degrees and the remaining field is 5 %; at pi standard depths the lag reaches a full 180 degrees, the signal is back in line with lift-off, and beyond it deeper flaws start plotting like shallower ones. Combined with the amplitude loss, this fixes the useful inspection depth at roughly three standard depths regardless of available gain.
Use this when building or checking a calibration curve, when deciding whether two reflectors will be separable, and when explaining why a known deep flaw plots where it does on the screen.
Worked example
| Material property entered as | rho |
| Electrical resistivity | 72 uohm.cm |
| Conductivity | 2.39 %IACS |
| Relative permeability | 1 |
| Test frequency | 100 kHz |
| Depth of flaw origin | 1 mm |
| Standard depth of penetration | 1.35 mm |
| Phase lag at that depth | 42.4 deg |
| Field strength at that depth | 47.7 % |
| Depth for 90 degrees lag | 2.12 mm |
| Depth for 180 degrees lag | 4.24 mm |
304 stainless (rho = 72 uohm.cm, mur = 1) at 100 kHz gives delta = 1000 x sqrt(7.2e-7 / (pi x 1e5 x 1.256637e-6)) = 1.3505 mm. A flaw originating 1.0 mm below the surface sits at x/delta = 1.0/1.3505 = 0.74048 standard depths, so the lag is 0.74048 rad x 57.2958 = 42.4 degrees and the field there is 100 x e^-0.74048 = 47.7 %. Quadrature (90 deg) is reached at 1.3505 x pi/2 = 2.12 mm and half-cycle (180 deg) at 1.3505 x pi = 4.24 mm.
Use at your own risk — verify before you act
These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.