Dead element tolerance
Elements fail. They fail as open circuits, as short circuits, and as elements that still fire but at reduced amplitude, and a probe accumulates them over its life. The element check is therefore part of periodic equipment verification, not something to be done only when a problem is suspected.
The generally applied acceptance is that no more than a quarter of the elements in the active aperture may be inactive, and that no two inactive elements may be adjacent. The first limit protects sensitivity, which falls roughly as 20·log10((n − d)/n). The second protects the beam shape: a single missing element is a small perturbation, but a gap of two or more punches a hole in the aperture and raises side lobes, which puts spurious energy at angles the focal law does not account for.
Losing elements at the ends of the aperture is worse than losing them in the middle, because the ends contribute the largest delays and therefore the most steering and focusing authority. A probe that passes the count but has failures clustered at one end will show asymmetric sensitivity across a sectorial sweep.
Whatever the general rule, the number that governs is the one in the written procedure and the equipment verification standard being worked to. Record the element map, not just a pass or fail.
Worked example
| Elements in the active aperture | 16 |
| Inactive elements in the aperture | 3 |
| Longest run of adjacent inactive elements | 1 |
| Elements on the probe | 64 |
| Inactive elements on the whole probe | 5 |
| Maximum inactive fraction allowed | 25 % |
| Maximum inactive elements allowed in the aperture | 4 |
| Inactive elements in the aperture | 18.75 % |
| Inactive elements on the whole probe | 7.81 % |
| Approximate sensitivity loss | 1.8 dB |
A quarter of 16 elements is 4, so up to 4 inactive elements are allowed and 3 is acceptable provided no two are adjacent. Three out of sixteen is 18.75 percent, and the aperture has fallen from 16 to 13 active elements, a loss of 20 × log10(13/16) = 1.80 dB. Across the whole probe, 5 of 64 is 7.81 percent.
Use at your own risk — verify before you act
These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.