Grating lobe free pitch
A phased array is a sampled aperture. Sampling a wavefront at intervals of p produces the wanted main lobe plus repeats of it at the spatial-frequency aliases, the grating lobes. They appear at angles satisfying sin θ_g = sin θ_s ± m λ/p. A grating lobe only exists as a real beam when that sine has magnitude of one or less; otherwise the alias is evanescent and stays trapped at the aperture.
Setting the first order alias exactly at grazing incidence gives the design rule p ≤ λ / (1 + sin θ_max). For an array that only fires straight ahead, a pitch of one wavelength is enough. To steer to 30 degrees the pitch must fall to about two thirds of a wavelength, and to steer to 60 degrees to about half a wavelength.
Grating lobes matter because they are almost as strong as the main lobe and they arrive at a different angle, so a reflector they illuminate is plotted at the wrong place. On a sectorial scan they show as a curved ghost that moves the opposite way to the real indication as the probe is moved.
For a probe on a wedge the sampling happens in the wedge, so the wavelength that controls grating lobes is the wedge wavelength (2730 m/s in Rexolite or Perspex), not the wavelength in steel. Wedge wavelengths are shorter, which is why wedge probes need finer pitch for the same steering range.
Compression or shear
Worked example
| Probe frequency | 5 MHz |
| Wave mode | shear |
| Material velocity | 3240 m/s |
| Required steering angle (half range) | 30 deg |
| Actual probe pitch | 0.6 mm |
| Wavelength | 0.648 mm |
| Maximum grating lobe free pitch | 0.432 mm |
| Grating lobe free steering for the actual pitch | 4.59 deg |
| Margin (p_max − actual pitch) | -0.168 mm |
λ = 3240/(5 × 1000) = 0.648 mm. sin 30° = 0.5, so p_max = 0.648/1.5 = 0.432 mm. A 0.6 mm pitch probe gives λ/p = 1.08, so grating lobe free steering only reaches asin(0.08) = 4.59°, and the margin is 0.432 − 0.600 = −0.168 mm.
Use at your own risk — verify before you act
These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.