Scan line count and index offsets

One pass of a phased array probe covers a band of material set by the electronic coverage width, or by the beam width where a single law is used. Covering a wider area means parallel passes, offset by less than the coverage so that adjacent passes overlap.

The index step is coverage × (1 − overlap) and the number of lines is ceil((required width − coverage) / index step) + 1. Overlap is not a luxury: encoder slip, probe wander and the tapering sensitivity at the edge of a beam all eat into the nominal coverage, and a nominally exact fit leaves unexamined strips.

Because the line count is a ceiling function, the last pass usually overlaps more than the others. That surplus is free coverage, and it is better placed at the end of the sequence than spread evenly, since it keeps the index positions on round numbers that are easy to set and easy to record.

Worked example

Width to be covered300 mm
Coverage width of one pass40 mm
Overlap between passes20 %
Index offset between passes32 mm
Number of scan lines10
Width actually covered328 mm
Coverage in excess of the requirement28 mm

With 40 mm of coverage and 20 percent overlap the index step is 40 × 0.8 = 32.00 mm. After the first pass 300 − 40 = 260 mm remains, needing ceil(260/32) = 9 further passes, so 10 lines in all. They span 9 × 32 + 40 = 328.00 mm, 28.00 mm more than required.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.