Dose rate at distance

The exposure rate from an unshielded gamma source is the product of two things: how many photons the source emits, set by its activity A, and how energetic and penetrating those photons are, captured by the specific gamma-ray constant Gamma. Spread over a sphere of radius d, the result is D = Gamma x A / d^2. With Gamma in R.m^2/(h.Ci), A in curies and d in metres, D comes out in roentgen per hour.

The roentgen is a unit of exposure in air. For photon energies used in industrial radiography, 1 R corresponds to about 8.76 mGy of air kerma, and for practical dose accounting the field convention is 1 R = 8.76 mSv. That gives 8760 uSv per roentgen, the conversion used here. Older survey meters read directly in mR/h, which is why both scales are shown.

This is the number that sets everything on the safety side of the job: the barrier distance, the stay time, whether a colleague can work at the far end of the bay, and how long a source can sit exposed before someone accumulates a reportable dose. It assumes a bare source in free air. A collimator, the projector body itself and any intervening steel all reduce the real rate, and scatter from nearby walls and the ground adds a few percent back. Always confirm with a survey meter – the calculation tells you where to stand while you take the first reading, it does not replace it.

Worked example

Isotopeir192
Present activity50 Ci
Distance from source5 m
Override gamma constant (0 = use table)0
Gamma constant used0.48
Dose rate8409.6 uSv/h
Exposure rate960 mR/h
Dose rate8.4096 mSv/h

D = 0.48 x 50 / 5^2 = 24 / 25 = 0.96 R/h. In mR/h that is 960. Converting at 8760 uSv per R: 0.96 x 8760 = 8409.6 uSv/h, i.e. 8.4096 mSv/h.

Use at your own risk — verify before you act

These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.