Lock-in thermography diffusion length

Lock-in (modulated) thermography heats the surface with a sinusoidal source and extracts the amplitude and phase of the surface temperature at the excitation frequency. The heat entering the part propagates as a heavily damped thermal wave whose amplitude decays by 1/e over the thermal diffusion length µ = √(α/(π·f)). That single expression sets the whole technique: depth is selected by choosing the modulation frequency, low frequencies reaching deep and high frequencies confining the interrogation to the near surface.

Because µ goes as the inverse square root of frequency, depth is expensive. Halving the frequency buys only a 41% increase in reach, and each halving doubles the acquisition time since several full cycles must be averaged. The amplitude image is usable to roughly one diffusion length; the phase image reaches deeper, conventionally about 1.5 times µ, and it has the decisive practical advantage of being largely immune to non-uniform heating, surface emissivity variation and reflections — the reasons phase images are what most lock-in procedures actually evaluate.

The thermal wavelength λ = 2πµ is the distance over which the thermal wave completes one cycle of phase. It is worth quoting alongside the diffusion length because depth is often estimated from the measured phase shift as a fraction of that wavelength, and because a defect deeper than about half a wavelength produces a phase change too small to separate from noise.

Run the calculation both ways. Given the frequency, it tells you how deep you are looking; given the deepest feature of interest, it gives the frequency to set. Then confirm on a reference standard of the same material and thickness containing known defects at known depths — the 1.5 µ phase-depth factor is a widely used approximation, not a physical constant, and the factor that applies to your material and defect type should be established by measurement.

Worked example

Materialcfrp
Thermal diffusivity (custom)0.4 mm2/s
Modulation frequency0.1 Hz
Phase depth factor1.5
Depth of interest3 mm
Cycles to average5
Diffusivity used0.4 mm2/s
Thermal diffusion length1.128 mm
Probing depth, phase image1.69 mm
Thermal wavelength7.09 mm
Frequency for the depth of interest0.0318 Hz

CFRP through-thickness at α = 0.40 mm²/s. µ = √(0.40/(π x 0.1)) = √1.27324 = 1.128 mm, so the phase image reaches about 1.5 x 1.128 = 1.69 mm and the thermal wavelength is 2π x 1.128 = 7.09 mm. To reach 3 mm the diffusion length must be 3/1.5 = 2 mm, needing f = 0.40/(π x 4) = 0.0318 Hz — about 31 s per cycle.

Use at your own risk — verify before you act

These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.