ToFD near-surface dead zone

The lateral wave is not an instantaneous spike; it occupies a time equal to the duration of the transmitted pulse. Any diffracted signal arriving inside that interval is buried in it. Setting the arrival time of a tip at depth d equal to the end of the lateral wave pulse and solving for d gives the shallowest depth at which a diffractor can be resolved: d = sqrt(S*c*tau + (c*tau/2)^2), where tau is the pulse duration and S is half the probe centre separation.

Everything shallower than that is the near-surface dead zone, and it is usually the limiting factor of a ToFD examination. It sits exactly where many fabrication and service defects occur, including lack of side-wall fusion at the cap, undercut and toe cracking. Its depth grows with both the probe centre separation and the pulse duration, so a heavily damped higher-frequency pair at the smallest workable PCS gives the smallest dead zone.

Pulse duration is taken as the number of cycles in the transmitted pulse divided by the centre frequency. Two cycles at 5 MHz is 0.4 microseconds. Measure it on the actual A-scan rather than assuming it: wedge damping, cabling and receiver bandwidth all change the effective pulse length, and because the dead zone grows with the square root of the pulse duration, a probe that rings for four cycles instead of two makes it about 40 percent deeper.

The dead zone cannot be removed by processing, so ISO 10863 and ASME require it to be covered by a complementary technique, normally a pulse-echo or phased-array scan of the near surface, or a second ToFD setup dedicated to the upper zone. Calculate it, state it in the procedure and record it on the report.

In practice the probe angle is chosen first and the PCS follows from it: with the beams crossed at two-thirds of the wall, PCS = (4/3)·T·tan θ. A 70° setup therefore carries a much wider PCS than a 45° one on the same wall, and since the dead zone grows with the probe separation, the chosen angle directly changes the dead zone. This calculator accepts either the angle (and derives the PCS) or a PCS you have already fixed.

Compression wave

Worked example

How to set the PCSderive
Refracted beam angle60 deg
Wall thickness25 mm
Probe centre separation70 mm
Probe centre frequency5 MHz
Cycles in the pulse2
Compression velocity5900 m/s
PCS in use57.7 mm
Half separation28.87 mm
Pulse duration0.4 µs
Pulse length in material2.36 mm
Near-surface dead zone depth8.34 mm

60° on 25 mm wall: PCS = (4/3)×25×tan60° = 57.74 mm, S = 28.87 mm. τ = 2/5 = 0.4 µs, pulse = 5.9×0.4 = 2.36 mm. d_dead = √(28.87×2.36 + 1.18²) = √69.52 = 8.34 mm. The same probes at 70° give PCS = 91.6 mm and a dead zone of 10.4 mm — the angle changes the answer.

Use at your own risk — verify before you act

These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.