ToFD probe selection by wall thickness
ToFD probe choice is a trade-off between resolution and penetration. Higher frequency and fewer cycles shorten the pulse, which shrinks both dead zones and improves depth resolution, but attenuation and beam divergence then limit how much wall a single setup can cover. Larger elements produce a narrower, more directional beam that carries further, at the cost of covering less depth range from one probe position.
Beam angle follows thickness. Thin sections use steep angles so that a workable probe centre separation still puts the crossing point at two-thirds of the wall; heavy sections use 45 to 60 degrees so the sound path, and therefore the attenuation, stays manageable. Above roughly 50 mm a single setup can no longer cover the wall and the examination is split into depth zones with a probe pair for each.
Because ToFD relies on the diffracted longitudinal arrival being first, the probes are compression-wave probes on angled wedges. Beam spread is used deliberately: a small element with a wide divergence illuminates a large depth range from a single position, which is why ToFD probes are usually much smaller than the pulse-echo probes used on the same joint.
The values here are banded starting points taken from the ToFD standards. They are a starting point only. The setup has to be proved on a reference block representative of the material, thickness, surface condition and weld geometry, and adjusted if the required flaw sensitivity, depth accuracy and near-surface coverage are not demonstrated.
f, θ, element size and number of setups are read from thickness bands λ = c / f d_cross = (2/3) × T PCS = 2 × d_cross × tan(θ) τ = n_cycles / f near-surface dead zone = √( S × c × τ + (c × τ / 2)² )
- Banded starting points only. The governing procedure and a demonstration on a representative reference block take precedence.
- Frequency bands are ranges in the standards; the value shown is a common working choice within the band. Raising the frequency shortens the pulse and shrinks the dead zones.
- The dead zone uses the cycle count entered (default two, a well damped probe). A probe ringing for four cycles makes it about 40 percent deeper, since the dead zone grows with the square root of the pulse duration.
- ISO 10863 requires the near-surface dead zone to be covered by a complementary technique whatever probe is chosen.
Reference: ISO 10863:2011 Tables 1 and 2 (recommended TOFD setups and probe parameters by wall thickness); ASTM E2373/E2373M. Values are banded starting points and must be verified against the current edition and the governing procedure.


