d = √((c · t / 2)² − S²)
Worked example PCS 70 mm, 15.1 µs, probe delay 2.5 µs d = 12.51 mm half path 37.17 mm
NDT calculators · TOFD
Probe separation, timing, depth, sizing and the dead zones of a ToFD set-up. Each of the 17 shows the governing formula, a worked example and the standard it comes from.
1 works without an account; the other 16 open with a free member account. Join free or sign in.
d = √((c · t / 2)² − S²)
Worked example PCS 70 mm, 15.1 µs, probe delay 2.5 µs d = 12.51 mm half path 37.17 mm
Probe separation, timing, depth, sizing and the dead zones of a ToFD set-up.
TOFD Depth and sizing
Converts a measured ToFD arrival time into the depth of the diffracting tip below the scanning surface.
ISO 10863 · ASME V Free
TOFD Dead zones and errors
Calculates the depth below the scanning surface within which diffracted signals are masked by the lateral wave pulse.
ISO 10863 · BS 7706
TOFD Timing, gates and scanning
Predicts where the lateral wave should appear on the ToFD A-scan for a given probe centre separation, velocity and probe delay.
ISO 10863 · ASTM E2373
TOFD Probes and PCS
Computes the beam divergence from the probe crystal diameter and frequency, and the band of wall depth a single ToFD setup actually insonifies at the weld centreline.
ISO 10863
TOFD Depth and sizing
Sizes a planar flaw by converting the upper and lower tip diffraction times to depths and taking the difference.
ISO 10863 · ISO 15626
TOFD Probes and PCS
Sets the probe centre separation so the two ToFD beam axes intersect at a chosen depth in the wall.
ISO 10863 · ASME V
TOFD Probes and PCS
Gives the standard single-setup ToFD probe centre separation for beams crossing at two-thirds of the wall thickness.
ISO 10863 · ASTM E2373
TOFD Timing, gates and scanning
Predicts the backwall reflection time on a ToFD A-scan and the time window between the lateral wave and the backwall.
ISO 10863 · ASME V
TOFD Timing, gates and scanning
Derives the digitising gate start, end and width from the predicted lateral wave and backwall arrival times plus operator margins.
ISO 10863 · ASME V
TOFD Probes and PCS
Splits a thick wall into ToFD depth zones and gives the PCS and arrival-time window for the selected zone.
ISO 10863 · ASME V
TOFD Probes and PCS
Solves the material compression velocity and the total wedge delay from the measured lateral wave and backwall arrival times on a known thickness.
ISO 10863 · ASME V
TOFD Dead zones and errors
Calculates how close to the far surface a diffracted signal can be resolved before it is masked by the backwall reflection.
ISO 10863 · BS 7706
TOFD Depth and sizing
Converts a timing uncertainty into a depth uncertainty at a stated depth, using the gradient of the ToFD time-to-depth curve.
ISO 10863
TOFD Timing, gates and scanning
Checks the ToFD scan increment against the allowed value and the encoder resolution, and gives the maximum scan speed for the pulse rate and averaging in use.
ISO 10863 · ASME V
TOFD Probes and PCS
Recommends ToFD probe frequency, beam angle, element size and number of setups for a given wall thickness, with the resulting PCS and near-surface dead zone.
ISO 10863 · ASTM E2373
TOFD Dead zones and errors
Shows how much depth error an error in the probe centre separation produces at a given flaw depth, and the PCS tolerance needed for a target depth accuracy.
ISO 10863
TOFD Dead zones and errors
Gives the apparent depth of a diffractor that is not on the centreline between the probes, for an offset across the weld and along the scan axis.
ISO 10863
Check the spelling, try a method code such as UT or RT, or clear the filters.
193 calculations with worked examples
955 discussions · 383 answers