ToFD probe delay and velocity from lateral wave and backwall

Wave mode: compression

The lateral wave and the backwall give two equations in the two quantities a ToFD setup cannot measure directly: the material compression velocity and the total delay through the two wedges. Both signals pass through the same wedges, so subtracting the two arrival times removes the delay and leaves t_BW - t_LW = (2*sqrt(S^2 + T^2) - PCS)/c, which gives the velocity. The delay then follows from the lateral wave alone, t_delay = t_LW - PCS/c.

This is the routine way to set up a ToFD depth scale. It needs only a piece of the parent material of known thickness, often the component itself away from the weld, and it calibrates velocity and delay in the same geometry, at the same temperature, with the same wedges and couplant that will be used for the scan. No side-drilled holes and no separate calibration block are required for depth linearity, although a reference block is still needed to demonstrate sensitivity.

Always sanity-check the result. A velocity far from about 5900 m/s in ferritic steel almost always means the PCS or the thickness used is wrong, not that the material is unusual. A delay that drifts between checks means wedge wear, a temperature change or a probe holder that has moved, and any depths measured in between are suspect. Repeat the check at the end of the scan and record both results.

The velocity solved here is the compression velocity, because both the lateral wave and the first backwall arrival are longitudinal. Entering shear-wave times or a shear velocity anywhere in this calculation is a category error and will produce nonsense depths.

S = PCS / 2
path difference = 2 × √(S² + T²) − PCS
c = path difference / (t_BW − t_LW)
t_delay = t_LW − PCS / c
delay per probe = t_delay / 2

Members-only calculator

Sign in or join free to use the full NDT toolbox.

Sign in View memberships

Notes:
  • Place both cursors on the same feature of each signal, normally the first zero crossing, so the delay cancels correctly.
  • The thickness must be measured independently, for example with a calibrated thickness gauge, not taken from the drawing.
  • Repeat the check at the end of the scan. A change in delay invalidates the depths recorded in between.
  • Delay per probe is only meaningful when the two wedges are the same design and equally worn.

Reference: ISO 10863:2011 §8 (setting the depth scale from lateral wave and backwall); ASME BPVC Section V, Article 4, Mandatory Appendix III

These calculators support — never replace — calculations against the governing code edition and your written procedure. Verify results independently before use.

Related calculators

193 calculations with worked examples

NDT calculators

All calculators

955 discussions · 382 answers

Latest forum discussions

Open the forum
Process columns and pipe racks against a blue sky

Looking for work?

105 live NDT vacancies worldwide

Upload your CV once and get new vacancies for your methods by email.

Sponsors of NDT Inspect
Magnaflux
Sponsor slot open