NDT calculators · ET

Eddy Current calculators.

Depth of penetration, phase lag, frequency choice, probes and tube wall loss. Each of the 12 shows the governing formula, a worked example and the standard it comes from.

1 works without an account; the other 11 open with a free member account. Join free or sign in.

Eddy current

12 calculators

Core calculations first. Pick a topic or search to narrow them down.

Eddy current 12

Depth of penetration, phase lag, frequency choice, probes and tube wall loss.

  • ET Frequency and depth

    Standard depth of penetration

    Eddy current standard depth of penetration (skin depth) from resistivity, relative permeability and test frequency, with field strength at the far surface.

    ASNT Free

  • ET Frequency and depth

    Eddy current phase lag with depth

    Phase lag of the eddy current at a given depth (57.3 degrees per standard depth of penetration), with the depths at which the lag reaches 90 and 180 degrees.

    ASNT

  • ET Probes and coils

    Fill factor and lift-off sensitivity

    Fill factor for a bobbin or encircling coil from coil and bore diameters, with radial clearance and the loss of coupling caused by additional clearance.

    ASME V

  • ET Tubing, scanning and conductivity

    Conductivity and resistivity conversion

    Convert between %IACS, conductivity in MS/m and resistivity in uohm.cm or nohm.m, correcting the reading to the 20 C IACS reference temperature (100 %IACS = 58.00 MS/m).

    ASTM E1004

  • ET Probes and coils

    Bobbin probe sizing for tube ID

    Bobbin probe diameter range for a tube, sized to a target fill factor band (commonly 85 to 90 %), with the radial clearance each choice leaves.

    ASME V

  • ET Frequency and depth

    Test frequency for a target depth

    Eddy current test frequency needed to place a target depth at a chosen number of standard depths of penetration, with the resulting phase lag and field strength.

    ASNT

  • ET Tubing, scanning and conductivity

    Tube wall loss from phase angle

    Percentage through-wall loss and remaining wall in a tube, read from the measured signal phase angle by interpolating between the two bracketing calibration reflectors.

    ASME V

  • ET Probes and coils

    Coil impedance and reactance

    Inductive reactance, total impedance, phase angle and Q of an eddy current probe coil at the test frequency, plus the coil and cable resonant frequency.

    ASNT

  • ET Probes and coils

    Eddy current array coverage and pitch

    Active coverage width of an eddy current array from element count, pitch and width, with pass count for a given scan width and the maximum speed the multiplexing rate allows.

    ASTM E3052

  • ET Tubing, scanning and conductivity

    Scan speed and sampling rate

    Maximum eddy current scan speed for a required number of samples across the smallest reportable flaw, with the sample spacing and samples per flaw at the planned speed.

    ASME V

  • ET Tubing, scanning and conductivity

    Remote field transition and probe spacing

    Exciter-to-detector spacing for remote field testing of ferromagnetic tube, with the double-transit wall phase lag and attenuation at the chosen frequency.

    ASTM E2096 · ASNT

  • ET Frequency and depth

    Limit frequency for rod and tube testing

    Forster limit (characteristic) frequency for an encircling coil on a rod or tube, the frequency ratio f/fg at the test frequency, and the wall-to-skin-depth ratio.

    ASNT · ASTM E309

Formulas are the standard engineering forms. Values taken from codes — IQI classes, allowable stress, dwell tables, acceptance limits — are stated with the edition they come from.

Use at your own risk — verify before you act

These calculators support, and never replace, the judgement of qualified NDT and engineering personnel. Results are provided as is, without warranty of any kind, express or implied, and must be independently verified against the governing code edition named in your contract before being used in any inspection, acceptance, rejection, radiation-safety or fitness-for-service decision. By using them you accept full responsibility for how the results are applied; NDT Inspect, its owners and contributors accept no liability for any loss, damage, injury or death arising from their use or from reliance on them. If a result matters to safety, check it by hand and have it reviewed by a competent person.