Bobbin probe sizing for tube ID
Probe selection for tube examination is a compromise between coupling and clearance, and fill factor is how it is expressed. Fill factor is the square of the diameter ratio, eta = (d_probe / D_id)^2, so the probe diameter that achieves a required fill factor is d_probe = D_id x sqrt(eta). Working the relationship in this direction is what a technician actually needs: the tube dimensions are given, and the question is which probe out of the kit to fit.
Too small a probe loses signal to the annular gap and, worse, loses noise margin. Wobble becomes a larger fraction of the clearance, small OD-initiated indications sink into the noise, and the phase readings that carry depth information get noisier along with everything else. Too large a probe gives excellent coupling right up to the moment it jams in a dent, an ovalised span, a deposit or a U-bend - and retrieving a stuck probe from an in-service heat exchanger costs far more than the few percent of amplitude that a smaller probe would have given away.
Common practice settles at 85 to 90 % fill for bobbin probes in straight tubing, and lower - often 75 to 80 % - for U-bend and severely fouled tubes where passage matters more than sensitivity. Note that fill factor changes far more slowly than diameter: over the whole 85 to 90 % band the probe diameter only moves about 3 %, so a kit with a sensible diameter increment covers a wide range of tube sizes.
Whatever probe is selected, sensitivity has to be proved and not assumed: the calibration standard is examined with that same probe, at the same frequencies, and the response from the calibration reflectors is what sets the gain. A probe change mid-job is a re-calibration, not an adjustment.
D_id = OD - 2 x wall d_probe = D_id x sqrt(fill factor) radial clearance = (D_id - d_probe) / 2 check: fill factor = (d_probe / D_id)^2
- Round the recommended diameter to a probe that actually exists in the kit, then re-check the fill factor rather than assuming it.
- Size on minimum wall where the tubing is bought to a minimum-wall specification - a thicker-than-nominal wall means a smaller bore than the drawing implies.
- Run the largest candidate probe through the worst tube in the bundle before committing the whole job to it.
- U-bend, deposit-fouled or dented tubing usually needs a lower fill factor and a correspondingly higher gain; qualify that combination on the calibration standard.
- A probe change is a re-calibration. Re-run the calibration standard with the probe that will be used.
Reference: ASME BPVC Section V, Article 8 - Eddy Current Examination of Tubular Products (probe selection to achieve the required calibration sensitivity); the 85 to 90 % fill range is common industry practice rather than a code number


