Helium test detectability

A mass spectrometer leak detector is calibrated in helium. Its quoted minimum detectable leak rate — say 1×10⁻⁷ mbar·L/s in sniffer mode — assumes the leak is emitting pure helium at the reference pressure difference, normally 1 bar. On real plant the system is rarely charged with pure helium at 1 bar: it is more often a 5–20% helium-in-nitrogen mix at several bar. What the detector sees is only the helium fraction of the total flow, so the smallest real leak that will trip the instrument is the instrument's own floor divided by the helium partial pressure expressed as a multiple of the reference differential.

Helium partial pressure is the concentration times the absolute pressure: 10% helium at 3 bar absolute is 0.3 bar of helium, three tenths of the reference differential, so the detectable leak is 1/0.3 = 3.3 times worse than the datasheet figure. Raising the test pressure buys sensitivity in the same proportion as raising the concentration, which is often the cheaper option when the item can take the pressure.

Run the calculation the other way to size the charge: the minimum helium concentration needed to prove a given acceptance limit is the instrument floor divided by the acceptance limit, scaled by the reference and test pressures. If that concentration is not achievable, the test cannot demonstrate the criterion however carefully it is performed, and the procedure has to change — a hood or vacuum-envelope technique, a lower instrument floor, or a different method entirely.

Two practical limits sit under the arithmetic. Atmospheric air already contains about 5.2 ppm of helium, and a poorly ventilated space with a recent helium spill contains far more, so the working background is usually well above the instrument's stated floor. And the response of a sniffer depends on probe standoff, scan speed and dwell — a demonstrated response to a calibrated leak on the actual item is worth more than any of these numbers.

p_He = concentration × absolute test pressure
sensitivity scale = p_He / reference differential
Q_min = detector MDL / scale
minimum concentration = 100 × MDL × p_ref / (Q_allow × p_test)

Members-only calculator

Sign in or join free to use the full NDT toolbox.

Sign in View memberships

Notes:
  • Air contains about 5.2 ppm helium as background; in a confined space after a helium release the working background is far higher and swamps the instrument floor.
  • Raising the absolute test pressure improves detectability in exact proportion, and is often cheaper than raising the helium concentration.
  • Confirm the calculated sensitivity by responding to a calibrated leak placed on the item, at the same probe standoff and scan speed used for the test.
  • The scaling assumes leak flow proportional to helium partial pressure — valid for molecular flow, approximate elsewhere.

Reference: General engineering — helium partial-pressure scaling per ASTM E499/E499M (detector probe), ASTM E498 (tracer probe) and ISO 20485:2017; instrument calibration against a standard leak per ISO 20486.

These calculators support — never replace — calculations against the governing code edition and your written procedure. Verify results independently before use.

Related calculators

193 calculations with worked examples

NDT calculators

All calculators

955 discussions · 382 answers

Latest forum discussions

Open the forum
Petrochemical plant with distillation columns

Join free

Join 5,341 NDT professionals for free

Groups, forum, resume upload, vacancy alerts and the member directory — no charge.

Sponsors of NDT Inspect
Magnaflux
Sponsor slot open