Pressure decay leak rate

A pressure change test measures the whole system at once: seal it at a known pressure, hold, and see how much pressure is lost. For a fixed volume the leak rate is the pV throughput Q = ΔP·V/Δt, so a 40 mbar loss from a 50 litre system over 10 minutes is 40 × 50 / 600 = 3.3 mbar·L/s. The method is cheap and needs no tracer gas, but it is blind to leak location and it is entirely at the mercy of temperature.

Temperature is the reason most pressure decay tests are wrong. Gas in a rigid volume obeys P/T = constant, so a fall of just 1 K from 293 K drops the pressure by about 0.34% — on a 6 bar absolute system that is 20 mbar of apparent loss with no leak at all. The correction is to refer the final pressure back to the starting temperature, P₂' = P₂ · T₁/T₂, using absolute pressures and absolute temperatures, and to take the leak-induced drop as P₁ − P₂'. Gauge pressures cannot be used in this correction. Where the correction is a large fraction of the raw drop, the test is not really measuring a leak; extend the hold, insulate the item, or move to a tracer-gas method.

The floor on sensitivity is set by the instrument, not the physics: the smallest leak the test can see is roughly the pressure resolution times the volume divided by the hold time. Buying sensitivity therefore means holding longer or measuring pressure better — increasing the volume makes it worse. Allow a thermal stabilisation period before starting the clock, since compression during pressurisation warms the gas and the subsequent cool-down looks exactly like a leak.

P₂' = P₂ × (T₁ + 273.15) / (T₂ + 273.15)
ΔP_corrected = P₁ − P₂'
Q = ΔP_corrected × V / Δt
Q_min ≈ gauge resolution × V / Δt

Members-only calculator

Sign in or join free to use the full NDT toolbox.

Sign in View memberships

Notes:
  • Absolute pressures and absolute temperatures only — the correction is invalid with gauge pressure.
  • Measure gas temperature, not ambient air: a surface-mounted probe on a thick-walled vessel lags the contents.
  • Allow the item to stabilise thermally after pressurising before starting the hold; compression heating mimics a leak as it decays.
  • A rising corrected pressure means the temperature measurement is wrong or the volume is not rigid.
  • The test finds a total leak rate only. It cannot locate the leak — follow up with bubble, tracer gas or ultrasonic methods.

Reference: ASME BPVC Section V, Article 10 — pressure change test appendix; ideal gas law temperature correction. Terminology per ISO 20484:2017.

These calculators support — never replace — calculations against the governing code edition and your written procedure. Verify results independently before use.

Related calculators

193 calculations with worked examples

NDT calculators

All calculators

955 discussions · 382 answers

Latest forum discussions

Open the forum
Refinery lit up at dusk, reflected in water

Free tools

193 NDT calculators with worked examples

Beam geometry, near field, TOFD depth, source decay, penetrant dwell, leak rates and more.

Sponsors of NDT Inspect
Magnaflux
Sponsor slot open