RMS (Rapid Motion Scanner) operating manual
-
DiscussionNDT-Inspect started the discussion Setup Parameters for HydroFORM probe (Corrosion Mapping) in the forum Corrosion Mapping4 years agoSetup Parameters & Written Instruction for HydroFORM probe (WT in excess of 25mm).
Probe: Auto-Detect or 7.5L64-I4
Wedge: Custom Definition for thick gasket plate:
Angle: 0º Orientation: Normal Height: 31mm Velocity: 1480.00m/s Primary Offset: -30.5mm Secondary Offset: 0.00mm
(saved in ‘User’ defined wedges as HydroFORM thk mtl)
Law: Linear at 0º with Overlap / 4 Element Aperture / First Element: 1 /
Last Element: 64 / 4mm focus depth
Pulser: Low Energy / Optimum PRF
Reciever: Bandpass Filter 7.7 MHz / Video Filter: ON / Points Qty: 640
Setup:
1) Attach Gasket Plate to Probe Housing. Place Probe assembly on V1 block and flood column with water. Agitate probe to free water column from bubbles and to ensure steady couplant flow.
2) Set display to A-S- (C-Scan left ‘OFF’).
3) Set Range Start to -5mm to show first interface echo.
4) Reduce Gain until Interface Echo approx. 100% FSH.
5) Set Range to see 2 backwall echoes.
6) Set Gate ‘I’: Start: -5mm / Width: To cover Interface Echo / Mode: ‘I/’
7) Set Gate ‘A’: Start: Adjacent to Interface Echo / Width: To cover 1st BWE / Mode: Synchro ‘/’
8) Set Gate ‘A’: Start: Adjacent to Gate A End / Width: To cover 2nd BWE / Mode: Synchro ‘/’
9) Select ‘Thickness’ tab in Gates/Alarms menu; Source: ‘A/ - I/’, set ‘Min’ to Gate A Start value, set Max to 2 x Nominal Wall thickness (this can be changed at analysis stage, but causes problems if set too low prior to inspection).
10) Select ‘Display’ menu. Activate C-Scan.
11) Select ‘View Settings’ submenu, set Source 1 to ‘Thickness’; Source 2 to ‘A%’
12) Select ‘Properties’ submenu. Select Colour Palette. Check ‘Thickness’ values are the same as entered in Step 9.
13) Select ‘Sizing’ Menu; ‘Curve Setup’. Build Manual TCG as follows:
Point 1: Position after interface, no TCG gain
Point 2: Position at 1st BWE, add TCG gain to bring 1st BWE to 100% FSH
Point 3: Position at 2nd BWE, add TCG gain to bring 2nd BWE to 100% FSH
Point n: Repeat above steps for subsequent points until TCG covers inspection
range.
14) Select ‘Scan’ > ‘Inspection’. Select ‘Encoder 1’. Calibrate encoder as normal procedure. Reference encoder resolution is 8 counts/mm, site cal gives 7.77 counts/mm.
15) Select ‘File’ > ‘Data Settings’. Confirm ‘Data Selection’ is for ‘All A&C Scans’, and define scan file save name. Note – adding a hastag ‘#’ in place of a numeral in the file name will cause the Omniscan to save subsequent files with consecutively-numbered file names for each punch of the ’SAVE’ button, reducing time between scans if so desired.
-
Setup Instructions for the MX2
Firstly connect the probe to the OmniScan. If making a double sided setup connect probes to splitter and then connect splitter to the OmniScan.
FILE > NEW
WIZARD > PART+WELD (Follow through wizard)
Enter THICKNESS
WELD TYPE = V OFFSET
Enter weld configuration
FINISH
WIZARD > SETUP (Follow through wizard)
MODIFY SETUP > NEXT
PROBE = If not auto recognised choose probe
WEDGE = Add wedge
NEXT
LAW CONFIG = SECTORIAL OR COMPUND (this will be in scan plan)
First Element = Scan Plan
NEXT
ANGLES = Enter MIN and MAX angles and ANGLE STEP (scan plan)
NEXT
INDEX OFFSET = Scan Plan
FOCUS DEPTH = Either 1.2 thickness or for thickness above 25mm, root 1 setup and cap 1 or Scan Plan
GENERATE
UT SETTINGS > GENERAL
RANGE = Change to 6 or 7 skips for calibration purposes.
PULSER > ENERGY = Scan Plan/Technique sheet
RECIEVER > FILTER = Change to bandpass filter to suit probe frequency (10MHz = 10MHz Bandpass)
WIZARD > CALIBRATION
MODE = SENSITIVITY
START > NEXT > NEXT
Put the probe on the block and scan a hole between ½ T & T.
Adjust gates to start just before selected hole and just after. It helps to set gate thresh hold low (between 0 & 5%)
Move probe back and forward so the hole signal goes across the S-Scan and the green line will begin to appear. If it isn’t showing across the whole range, then add dB. Once the green line is maximized hit CALIBRATE.
Now move probe back and forward to get the green line to 80%. If it is to low or to high in places press CALIBRATE again. When it is good >
ACCEPT
FILE > SETUP > SAVE
Now take out 5 or 6 dB on the main gain setting
WIZARD > CALIBRATION
MODE=WEDGE DELAY
Select a hole between 1/2T & 2T
NEXT
DEPTH=Depth to centre of hole (look at drawing)
TOLERENCE=0.1
NEXT > NEXT
Set the gate to just before and just after the hole you have chosen.
THRESHOLD=20%
Get hole signal around 80% and move probe back and forward to get the white line across the screen.
CALIBRATE
Now check calibration by moving probe back and forward again. (you may have to change the gate now that the wedge delay has changed)
If calibration is bad, then repeat these steps.
If calibration is good, then ACCEPT.
WIZARD > CLAIBRATION > TYPE =SIZING
MODE=TCG
START > NEXT > NEXT > NEXT
Adjust range so hole can be picked up across all angles.
NEXT
Adjust gate over 1st hole
NEXT
Reduce dB so green line is around 50% and move probe back and forward to complete green line across the screen.
ADD POINT – now check the green line sits at 80% by moving probe back and forward. If bad CANCEL POINT and start that hole again.
If good, NEXT POINT
Repeat these steps for all TCG holes. Minimum of 3 within the test range.
When all holes complete ACCEPT TCG
UT SETTINGS > RANGE
Adjust range to cover inspection area
GATES
Adjust gates to start above root area and below full skip / 2T
SAVE SETUP
-
-
Виды и методы неразрушающего контроля
-
ملخص الفحوصات اللااتافية / مقدمة باللغة العربية للمبتدئين وتنشيط لاصحاب الخبرات
Nondestructive testing in summary - Introduction (Arabic) - YouTube
-
Tube Inspection TechnicianProzscan Marine Offshore & NDT services · Qatar
Required Tube Inspection technician for upcoming shutdown in Qatar.
Academic Qualification: Any Degree/Diploma
Technical qualification: ASNT Level II in ECT, RFT and required experience in IRIS
Experience: […] -
-
-
-
-
Detection, characterization and sizing of hydrogen induced cracking in pressure vessels using phased array ultrasonic data processing
-
-
-
-
-
-
-
-
-
-
-
-
-